UV-VISIBLE
DATA COLLECTION FOR Fe(II) AND TOTAL Fe
by
ThaoNguyen Nguyen
(Laboratory
of Joseph W. Stucki, University of Illinois)
- Install
the Routine Sample Analyzer (RSA) into the Cary 5
Spectrophotometer
- Start the
Cary V computer program by clicking on the “CaryWinUV” icon on the desktop
- Select
the “Concentration” sub
program.
- Turn on
the Spectrophotometer and allow to warm up for 20 minutes.
- Select “File”, “Open Method”, then choose YOUR folder (user defined)
- Select “Set Up”, followed by:
* “Choose standards” to set the
number of standards -- note that sometimes blanks don’t count as standards.
* “Choose
samples” to set the number of samples
- Select “File”, “Save Method As”, then save into YOUR folder on Drive C
- Select “Zero It”
- Select “Rinse” (which appears on a pop-up
window) one time to drain the “Sampling Tube”
line, then immerse the “Sampling
Tube” into Dionized Water and rinse again at least 5 times, draining the
tube between each rinse
- Immerse
the “Sampling Tube” into Deionized Water, then
select “Fill and Read” as your
primary reading. (The reading for
water should be ZERO, or at least close to Zero).
- Select “Start”, followed by “Save as”. This time save data in a file named
“YOUR data” (user defined) in YOUR
folder on Drive D
- Repeat steps
10 and 11 for each sample.
HINTS:
- Between
each reading of different samples, Rinse
the “Sampling Tube” by draining first, then refilling with Deionized Water to remove any trace left by the old
sample.
- RINSE the
“Sampling Tube” TWICE with the
sample solution before taking the final reading in order to minimize
errors. Make sure that the
“Sampling Tube” is fully immersed into solution
VERY IMPORTANT:
Upon completion of all samples, RINSE the “Sampling Tube” with Deionized Water for AT LEAST 20 times before turning the
equipment off.
15. Perform
calculations using Fe analysis data
processing spreadsheet.