Use Of Mettler-Toledo
Redox Electrode
by
Christian Bogdal
(Laboratory of Joseph W. Stucki,
1. Read
the potential of a certified buffer solution
Manufacturer: |
Mettler
Toledo GmbH (www.mt.com) |
Product
designation: |
Redox
Buffer Solution 468 mV (pH 0.1) / 9883 |
Order
Number: |
51
340 066 |
This
value is called: U(read buffer)
2. The
internal reference of the electrode is calculated by the following formula:
U(int ref calc) = U(buffer
theo) – U(read buffer) .
In
this formula, U(buffer theo)
stands for the potential of the buffer relative to hydrogen. The values of U(buffer theo)
are temperature dependent and listed in table 1. Hence, they have to
be chosen according to the measured room temperature.
3. Verify
the accuracy of the electrode by:
|U(int ref calc) - U(int ref theo)| < ± 10
mV.
U(int ref theo) symbolizes the theoretical internal reference of the
electrode. This value also depends on temperature and is given in table 1.
4. If
the electrode presents satisfactory results for its internal reference (i.e., the
test of step 3 is valid), measure the potential of the sample solution, which is
recorded as U(read solution).
5. The
redox potential of the sample solution is obtained as
follows:
U(solution) = U(read
solution) + U(int ref calc) .
Table 1
|
Electr.Ag/AgCl |
Buffer 230 mV |
Buffer 470 mV |
T [°C] |
U(int ref theo) [mV] |
U(buffer theo) [mV] |
U(buffer theo) [mV] |
10 |
217.4 |
471.4 |
669.0 |
11 |
216.7 |
469.1 |
669.4 |
12 |
216.0 |
466.8 |
669.8 |
13 |
215.4 |
464.6 |
670.2 |
14 |
214.7 |
462.3 |
670.6 |
15 |
214.0 |
460.0 |
671.0 |
16 |
213.3 |
457.7 |
671.4 |
17 |
212.6 |
455.4 |
671.8 |
18 |
211.9 |
453.1 |
672.2 |
19 |
211.2 |
450.8 |
672.6 |
20 |
210.5 |
448.5 |
673.0 |
21 |
209.8 |
446.2 |
673.4 |
22 |
209.1 |
443.9 |
673.8 |
23 |
208.4 |
441.6 |
674.2 |
24 |
207.7 |
439.3 |
674.6 |
25 |
207.0 |
437.0 |
675.0 |
26 |
206.3 |
434.7 |
675.4 |
27 |
205.6 |
432.4 |
675.8 |
28 |
204.8 |
430.0 |
676.2 |
29 |
204.1 |
427.7 |
676.6 |
30 |
203.4 |
425.4 |
677.0 |
Instructions And Hints To
Optimize The Method, Increase Accuracy, And Handle The Electrode Appropriately
1.
Rinse the electrode between each sample with distilled water (deoxygenated
distilled water if working in the glove box).
2.
Choose a buffer standard whose value U(buffer
theo) is as close as possible to the
expected redox potential of the sample U(solution).
3.
The electrode is equipped with a moveable “safe-lock” device. The refilling
port is sealed when the hole is visible trough the transparent part of the
safe-lock. When the hole is not visible, the electrode is ready to use. For
storage the refilling port is closed by turning the moveable safe-lock. For
measurements, remove the wetting cap and rinse with distilled water.
4.
For storage, rinse or clean the electrode after use and replace the wetting
cap, partly filled (1/3) with the reference electrolyte.
5. For refilling: Slip the safe lock approximately 20 mm
downwards. Refill the electrode with the recommended reference electrolyte, if
necessary. The level of a correctly refilled electrode is approximately 10 mm
below the filling hole.
6. Reference
electrolyte: KCl 3 M.
7. In
case of a slow response/drift check the following actions:
clean
and remove deposits from the metal ring with toothpaste or 2 M HCl
rinse
with distilled water
change
reference electrolyte
clean
diaphragm (as described below)
8. The
cleaning of the diaphragm involves different steps:
use
cleaning solution Pepsin/HCl (to remove proteins)
use
cleaning solution Thiourea (to remove sulfides)
rinse with
acetone or ethanol (for contamination by oily/organic liquids)
rinse
with 0.1 M HCl or 0.1 M NaOH
(for contamination by acid/alkaline soluble compounds)