1,10-PHENANTHROLINE
METHOD FOR Fe(II) AND Fe(III)
28 October, 200213
June 2005
6. 10 % hydroxiylamine
hydrochloride solution in water
6.7.
Ferrous ammonium sulfate (Fe(NH4)2(SO4)2.6H2O)
salt for standards
7.8.
High-purity water (preferably with a resistance of
18 megohm or more) for all dilutions and solutions
a. Fill the Dosimat bottle with
1 % Na-citrate solution.
b.a.On the Dosimat keyboard,
press <recall> 1, then press <Go>. Dil 2 2.000 mL should appear on the Dosimar display.
c.b.
Clean the pipet tip with water and wipe dry with a ChimwipeChimwipe.
d.c.
Fill a 50-mL beaker with water for rinsing
and cleaning the pipet tip between samples.
e.d.
Insert the pipet tip into the water, press the pipette
button to pipet the 2 mL of water. <Go>. When Dil 2 appears on
the Dosimat display, mMove
the pipet tip over a discard beaker and press <Go>the button
once more. The entire solution (about 22 mL) should then be expelled into
the beaker.
f.e.
Rinse and wipe the tip.
g.f.
Be sure Dil 1
2.0000 mL is displayed, then place pipet tip into the first digestion tube.
Press <Go>the pipette
button. . When
Dil 2 appears on the display, move
the pipet tip to the first flask and press the pipette
button<Go>
again. Cover the flask with parafilm. Repeat this process, transferring another
aliquot from the first digestion tube to the second flask. Rinse pipet tip and
dry with Chimwipe.
h.g.
Repeat step g f for
each digestion tube.
h. To prepare the total iron
analysis it is necessary to change the dilution
volume to 19 mL. On the keyboard press volume
1 + 9 and enter. The final volume must be
21 mL. With the Eppendorf pipette puttransfer 1 mL
of 10 % hydroxylamine
hydrochloride in the
flasks
i.
Cover ap
the glass flasks with parafilm and leave them for
24 h (or overnight) in the darkroom to complete reduction.
21.1.Replace the parafilm covers
on each flask, then place all flasks on the glass shelf above the Hg vapor lamp
in room W-316. Turn on the lamp and leave for 2 hrs.
22.21.
MRe-measure the
absorbance at 510 nm for the total Fe flasks.
23.22.
Calculate results as per UV-Visible Data Collection procedures.
a.Turn on vacuum pump for the
sipper system. Check to be sure holding bottle has sufficient room for more
waste solution.
b.a.Set the
wavelenght to 510 nm with the Tungsten lamp on and the mirror in either
the automatic or Tungsten position.
c.a.With water
in both the sample and reference beams, adjust the slits to 0.2 bandwidth.
d.a.Close the
sample shutter (R open, S closed) and position the span control to Calibrate.
Calibrate the instrument (display reading of 0000) using the calibrate knob to
the left of the absorbance plug-in module.
e.a.Change
shutter positions back to (S & R open) and move the span knob to 2. Zero
the instrument (display reading 0000) using the Zero Adjust knob (not labeled)
on the top right.
f.a.Set up the
computer and method for sampling according to the procedure UV-Visible Data Collection (it this
book).